15-parameters type is also available. ( 15 Standard curves can be memorized)
Current model is 5-parameters type.
EXTERNAL INTERFACE
Printer interface and RS-232C are built in.
MEMORY AND OTHER FUNCTIONS
Microprocessor enabled various functions such as the measurement of correlation curve, timer mode and
memory of standard data etc. as standard functions.
AUTO-ZERO FUNCTION
T=0% and T=100% can be adjusted with one touch of a button.
SMALL SAMPLE
Minimum sample volume is 1.0mL.
LONG LAMP LIFE
Long life span of lamp (2,000 hours)
EXCELLENT STABILITY
Excellent stability even in the low wavelength ranges.
ROUND OR SQUARE CUVETTE USABLE
Either round cuvette or square cuvette can be used without adapter.
FLEXIBLE POWER SOURCE
Power requirement is flexible, from 100 to 240VAC
EASY LAMP REPLACEMENT
Exchange of lamp is simplicity itself.
SPECIFICATIONS:
Wavelength Range
340 to 1000 nm
Wavelength Width
8nm in the entire region
Wavelength Accuracy
±2nm at 430 and 585nm
Wavelength Scale
1nm scale
Optical path length
10mm
Spectroscope
Diffraction grating, number of grooves 1200 lines/mm
Stray Radiant Energy
<0.5%T typical at 400nm with appropriate stray light filter
Display
LCD Display
Photometric Range
-0.200 to 3.000ABS
0.0 to 200% T
0.000 to 2000C
Photometric Accuracy
±2%T
Measuring mode
1. Normal mode
Timer mode – Measuring interval 1-60sec, Time limit 5940sec